In short
A pressure gauge is calibrated by applying known pressures and comparing its readings with a reference that is several times more accurate: either a deadweight tester, which makes the pressure with weights on a piston, or a calibrated reference gauge on a comparator pump. The gauge is read at several points up the scale and back down, the errors are recorded "as found", it is adjusted if allowed, and recorded again "as left" on a certificate traceable to national standards.
Two ways to make a known pressure
A. Comparator: 1. a pump or screw press makes the pressure; 2. a calibrated reference gauge reads it; 3. the gauge under test reads the same pressure on the second port; 4. a release valve lets the pressure off.
B. Deadweight tester: 5. calibrated masses on a piston of known area set the pressure (p = F ÷ A); 6. the piston floats when the pressure balances the weight; 3. the gauge under test reads it.

Deadweight tester or comparator
| Deadweight tester | Comparator with reference gauge | |
|---|---|---|
| How the pressure is known | From the masses and the piston area | Read on a calibrated reference gauge |
| Uncertainty (Fluke's figures) | As low as 0.008% of reading (0.015% without software corrections) | About 0.02% of scale for a good reference gauge and pump |
| Speed | Slower: masses are loaded for each point | Faster: pump to each point and read |
| Corrections | Local gravity, temperature, height | Settling time; check for leaks |
| Portability | A bench instrument | Can be carried; some work with one hand |
The steps
Labs differ in the detail; a typical sequence is:
- Check and prepare. Inspect the gauge, check its zero with no pressure, and let it reach the room's temperature. Mount it as it is used.
- Exercise it. Labs usually take it to full scale and back before reading, so the element settles.
- Read up the scale. Apply pressure in steps, typically five or more points spread over the range, and record the reference and the gauge's reading at each.
- Read back down. Repeat on falling pressure: the difference between rising and falling readings is the hysteresis.
- Record as found. These errors show how the gauge was performing when it came in.
- Adjust if allowed, and record as left. After any adjustment, repeat the readings.
- Certificate. The results, the reference used and its traceability, the uncertainty and the date go on the certificate. See what a calibration certificate shows.
How accurate must the reference be?
Clearly more accurate than the gauge: the traditional rule is a test uncertainty ratio of at least 4:1, the reference's uncertainty four times smaller than the gauge's tolerance. A 1% gauge needs a reference of 0.25% or better. See test gauges.
Calibrating a digital pressure gauge
The same way: known pressures, readings up and down, as found and as left. A digital gauge has no pointer to read between marks, so the readings are exact numbers; check its zero and battery first, and set it to the unit on the certificate.
Can you check a gauge yourself?
You can compare it with a calibrated test gauge on the same pressure, which shows whether it is close. A calibration that counts as evidence, for reports and audits, needs a reference traceable to national standards and a certificate. Check what your standard or client asks for.
Pressure gauge calibration at Staht
These are the general methods. Staht calibrates pressure gauges up to 1000 bar, Staht's G1 Pro and other makes, and each calibration comes with a UKAS traceable certificate, printed and digital, recording how the gauge performed on the day. Ask us how yours will be calibrated: see pressure gauge calibration.
The Staht G1 Pro is a calibrated digital Bluetooth pressure gauge for hydraulic pumps and cylinders, reading up to 1000 bar (14,504 psi), accurate to 0.25% of full scale (±2.5 bar, ±36 psi) and UKAS traceable. Its colour screen shows the pressure in bar, live and peak, and the free Staht Connect app shows the force in kN or lbf from the cylinder's piston area. £585.00 ex VAT (£702.00 inc VAT). See all our digital pressure gauges.
Questions
Known pressures are applied, from a deadweight tester or a pump with a calibrated reference gauge, and the gauge's readings are compared with the reference at several points up and down its scale, before and after any adjustment.
Calibrated masses sit on a piston of known area. Oil is pumped in until the piston floats; the pressure is then the weight divided by the piston area, and the gauge under test reads it.
A pump or screw press with two ports: one for a calibrated reference gauge and one for the gauge under test. Both read the same pressure, so the gauge is compared with the reference.
Typically five or more spread across the range, both rising and falling, so errors and hysteresis show over the whole scale. Your standard or the lab's procedure sets the exact number.
Check its zero with no pressure, then compare it with a calibrated test gauge at the pressures you use. For evidence, have it calibrated with a traceable certificate.
At least once a year, or as your standard or client asks, and after any drop, overload or doubtful reading. See how often to calibrate a pressure gauge.
Sources
- Pressure comparators and digital reference pressure gauges vs deadweight testers, Fluke
- EURAMET Calibration Guide No. 17: Guidelines on the calibration of electromechanical and mechanical manometers (v4.1, 2022), EURAMET
- What are ASME accuracy grades for mechanical pressure gauges?, Ashcroft
- Pressure gauge calibration, Staht
- Staht G1 Pro gauge user manual, Revision 1, 04/02/2026, Staht
Get a pressure gauge calibrated
Gauges up to 1000 bar, Staht's and other makes, each with a UKAS traceable certificate.