Guide

How often should a pressure gauge be calibrated?

Staht G1 Pro gauge in its open carry case, shown with a G1/4 quick connect fitting

In short

At least once a year, unless the gauge's maker, your standard or your client asks for more often, and again straight away after a drop, an overload or any reading you doubt. Site tests to BS 8539 and NASC TG4:25 need the test gauge calibrated within the last 12 months. Keep the certificate with the gauge, and put the calibration date in every test report.

Key figures

Usual intervalAt least once a year
Site tests to BS 8539 and NASC TG4:25Calibrated within the last 12 months
Calibrate again straight awayAfter a drop, an overload or a doubtful reading
Staht G1 ProCalibrated for one year from purchase; Staht recommends calibration every year

What decides the interval?

  • The maker's recommendation. Staht recommends calibrating the G1 Pro every year.
  • Your standard, specification or client. Some set a maximum interval, such as the 12 months for site tests to BS 8539 and NASC TG4:25.
  • How hard the gauge works. Heavy use, rough handling, vibration, pressure spikes and harsh conditions are reasons to calibrate more often.
  • Its history. If a gauge keeps coming back out of tolerance, shorten the interval.

Calibrate it again, whatever the date, after a drop, after it has been taken above its maximum pressure, or if it reads oddly, for example a dial needle that won't settle back to zero.

How is a pressure gauge calibrated?

The gauge is connected to a pressure source with a reference instrument, such as a pressure calibrator, a reference gauge or a deadweight tester, that is far more accurate than the gauge itself. The pressure is set to several points across the range, usually going up and coming back down, and at each one the gauge's reading is recorded beside the reference value.

The differences show whether the gauge is within its stated accuracy. If it isn't, it may be adjusted and checked again, or repaired or replaced.

What does a calibration certificate show?

A certificate records how the gauge read on the day it was calibrated. A good one names:

  • the gauge: make, model, serial number and range;
  • the laboratory, and who authorised the certificate;
  • the date of calibration, and often the date it is next due;
  • each test point: the reference pressure, the gauge's reading and the difference;
  • the uncertainty of the measurement;
  • how the reference equipment is traceable to national standards.

The international standard for calibration laboratories, ISO/IEC 17025, sets out what a laboratory's certificates must report. The certificate shows how the gauge read; it doesn't approve any test result made with it.

For the G1 Pro, the calibration date and the next calibration date are on its certificate, on its screen when it starts and in the Staht Connect app and web dashboard.

UKAS traceable or UKAS accredited?

  • UKAS traceable: the laboratory's reference equipment was itself calibrated, in an unbroken chain, back to the UK's national measurement standards.
  • UKAS accredited: the laboratory itself has been assessed by UKAS for that calibration, and its certificate carries the UKAS mark.

Check which your standard or client asks for before you book. Staht's pressure gauge calibration comes with a UKAS traceable certificate.

Pressure gauge calibration at Staht

Staht calibrates pressure gauges up to 1000 bar, its own G1 Pro and other makes, with a UKAS traceable certificate. For a price, contact us with the gauge's make, model and range, and tell us when you need it back. See pressure gauge calibration.

Questions

Sources

  1. Staht G1 Pro gauge user manual, Revision 1, 04/02/2026, Staht
  2. ISO/IEC 17025 General requirements for the competence of testing and calibration laboratories, ISO
  3. EURAMET Calibration Guide No. 17: Guidelines on the calibration of electromechanical and mechanical manometers (v4.1, 2022), EURAMET
  4. Pressure gauge calibration, Staht
  5. Pull tester gauge accuracy and calibration intervals, Staht