Pressure gauge accuracy explained

In short
Most pressure gauges state accuracy as a percentage of full scale (FS): the possible error is that percentage of the top of the scale, the same at every reading. A 1000 bar gauge accurate to 0.25% FS is within ±2.5 bar (±36 psi) anywhere on its scale, which is 0.25% of a 1000 bar reading but 2.5% of a 100 bar one. An accuracy stated as a percentage of the reading shrinks with the reading. So work out the error at the pressures you'll actually read.
Full scale or reading?
- % of full scale (FS), sometimes "of span": error = accuracy × full scale. Fixed in bar at every reading, so it is a bigger share of a low reading.
- % of reading: error = accuracy × the reading. It shrinks as the reading falls, so it stays the same share of the reading.
Take two gauges, both "0.25%". A 1000 bar gauge at 0.25% FS reading 100 bar is within ±2.5 bar. A gauge at 0.25% of reading, at the same 100 bar, is within ±0.25 bar. At 1000 bar both are within ±2.5 bar. Where a gauge doesn't start at zero, "span" is the range from the bottom of the scale to the top.
The G1 Pro at 0.25% of its 1000 bar full scale
| Reading | Possible error | As a share of the reading |
|---|---|---|
| 100 bar (1,450 psi) | ±2.5 bar (±36 psi) | 2.5% |
| 200 bar (2,901 psi) | ±2.5 bar (±36 psi) | 1.25% |
| 350 bar (5,076 psi) | ±2.5 bar (±36 psi) | 0.71% |
| 500 bar (7,252 psi) | ±2.5 bar (±36 psi) | 0.5% |
| 700 bar (10,153 psi) | ±2.5 bar (±36 psi) | 0.36% |
| 1000 bar (14,504 psi) | ±2.5 bar (±36 psi) | 0.25% |
Gauge accuracy calculator
Accuracy classes (EN 837-1)
In Europe, dial (Bourdon tube) gauges are made to EN 837-1, which gives each gauge an accuracy class: 0.1, 0.25, 0.6, 1, 1.6, 2.5 or 4. The class is the permitted error as a percentage of the span. A class 1.6 gauge reading 0 to 600 bar may be out by ±9.6 bar anywhere on its scale; a class 0.6 gauge on the same range by ±3.6 bar.
The main accuracy grades (ASME B40.100, used in the US)
| Grade | Permitted error, % of span |
|---|---|
| 4A | ±0.1% |
| 3A | ±0.25% |
| 2A | ±0.5% |
| 1A | ±1% |
| A | ±2% in the first and last quarters of the scale, ±1% in the middle half |
| B | ±3% in the first and last quarters of the scale, ±2% in the middle half |
Grade A is often written "2-1-2" and grade B "3-2-3". A 10,000 psi grade A gauge may be out by ±100 psi in the middle half of its scale and ±200 psi near either end.
Choosing the accuracy you need
- Find the limit your job sets. A standard, a specification or a client may give one. For example, BS 8539 site tests ask for a gauge reading within 10% of the true load: see pull tester gauge accuracy.
- Work out the error at your lowest reading, not at full scale. Use the calculator above.
- Choose the range to suit. A gauge whose full scale is far above your readings carries a large error at the bottom. For dial gauges, makers following ASME B40.100 advise working between a quarter and three quarters of the scale.
- Check what else the data sheet states, such as an error per degree away from the reference temperature.
Test gauges and calibration
When one gauge is checked against another, the reference must be clearly more accurate than the gauge under test. A test uncertainty ratio of 4:1 is the usual aim: to check a gauge accurate to ±1% of a range, use a reference within ±0.25% of it.
Accuracy is only true while the gauge stays calibrated. A drop, an overload or plain age can shift it, which is why gauges are calibrated at least once a year. See how often to calibrate a pressure gauge and Staht's pressure gauge calibration.
Accuracy as force
On a hydraulic cylinder, the gauge's error in bar becomes an error in force. Multiply it by the cylinder's effective area: on a 25.1 cm² cylinder, ±2.5 bar is ±0.63 kN at any load. See how gauge accuracy carries into force.
Questions
The possible error is 0.25% of the top of the gauge's scale, at every reading. On a 1000 bar gauge that is ±2.5 bar (±36 psi), whether it reads 100 bar or 900 bar.
For an accuracy of full scale: error = accuracy % × full scale ÷ 100. Divide the error by the reading, and multiply by 100, to get it as a share of the reading. The calculator on this page does both.
A full-scale accuracy is the same error in bar at every reading, so it is a larger share of a low reading. A reading accuracy is a share of the reading itself, so it shrinks as the reading falls.
A gauge made to EN 837-1 whose permitted error is 1.6% of its span. On a 0 to 600 bar gauge that is ±9.6 bar at any reading.
An ASME B40.100 grade: ±1% of span in the middle half of the scale and ±2% in the first and last quarters, written "2-1-2".
0.25% of its 1000 bar full scale: ±2.5 bar (±36 psi) at any reading. It is UKAS traceable and calibrated for one year from purchase.
Clearly more accurate than the gauge it checks. A test uncertainty ratio of 4:1 is the usual aim.
Sources
- BS EN 837-1 Pressure gauges. Bourdon tube pressure gauges. Dimensions, metrology, requirements and testing, BSI
- About pressure gauge calibration: accuracy classes, % of span, Calibrate.co.uk
- What are ASME accuracy grades for mechanical pressure gauges?, Ashcroft
- How to select a pressure gauge range, Ashcroft
- What is the accuracy/uncertainty of the device to be tested?, Fluke
- Staht Heavy Duty Test Kit brochure (the G1 Pro's range and accuracy), Staht
- Staht G1 Pro gauge user manual, Revision 1, 04/02/2026, Staht
A gauge accurate to 0.25% of full scale
The G1 Pro: up to 1000 bar (14,504 psi), ±2.5 bar at any reading, UKAS traceable. £585.00 ex VAT (£702.00 inc VAT).