Pressure gauge accuracy explained

Staht G1 Pro digital hydraulic pressure gauge from the front, its colour screen showing the pressure in bar with a trace

In short

Most pressure gauges state accuracy as a percentage of full scale (FS): the possible error is that percentage of the top of the scale, the same at every reading. A 1000 bar gauge accurate to 0.25% FS is within ±2.5 bar (±36 psi) anywhere on its scale, which is 0.25% of a 1000 bar reading but 2.5% of a 100 bar one. An accuracy stated as a percentage of the reading shrinks with the reading. So work out the error at the pressures you'll actually read.

Full scale or reading?

  • % of full scale (FS), sometimes "of span": error = accuracy × full scale. Fixed in bar at every reading, so it is a bigger share of a low reading.
  • % of reading: error = accuracy × the reading. It shrinks as the reading falls, so it stays the same share of the reading.

Take two gauges, both "0.25%". A 1000 bar gauge at 0.25% FS reading 100 bar is within ±2.5 bar. A gauge at 0.25% of reading, at the same 100 bar, is within ±0.25 bar. At 1000 bar both are within ±2.5 bar. Where a gauge doesn't start at zero, "span" is the range from the bottom of the scale to the top.

The G1 Pro at 0.25% of its 1000 bar full scale

ReadingPossible errorAs a share of the reading
100 bar (1,450 psi)±2.5 bar (±36 psi)2.5%
200 bar (2,901 psi)±2.5 bar (±36 psi)1.25%
350 bar (5,076 psi)±2.5 bar (±36 psi)0.71%
500 bar (7,252 psi)±2.5 bar (±36 psi)0.5%
700 bar (10,153 psi)±2.5 bar (±36 psi)0.36%
1000 bar (14,504 psi)±2.5 bar (±36 psi)0.25%
Pressure gauges

Gauge accuracy calculator

The top of the gauge's scale, in any pressure unit.

From the gauge's data sheet or certificate.

Most gauges state accuracy as a percentage of full scale (FS).

The pressure the gauge shows, in the same unit as the full scale.

Results

Possible error, ±2.500In the same unit as the full scale and the reading.
The true pressure is at least197.500
…and at most202.500
The error as a share of the reading1.25 %

For an accuracy stated as a percentage of full scale, the possible error is the same at every reading: error = accuracy % × full scale ÷ 100. For one stated as a percentage of the reading, it shrinks with the reading.

The figures shown when the page opens are an example: the Staht G1 Pro, 0.25% of its 1000 bar full scale, reading 200 bar.

The calculator covers the stated accuracy only. Check the data sheet for anything stated on top of it, such as the effect of temperature.

Accuracy classes (EN 837-1)

In Europe, dial (Bourdon tube) gauges are made to EN 837-1, which gives each gauge an accuracy class: 0.1, 0.25, 0.6, 1, 1.6, 2.5 or 4. The class is the permitted error as a percentage of the span. A class 1.6 gauge reading 0 to 600 bar may be out by ±9.6 bar anywhere on its scale; a class 0.6 gauge on the same range by ±3.6 bar.

The main accuracy grades (ASME B40.100, used in the US)

GradePermitted error, % of span
4A±0.1%
3A±0.25%
2A±0.5%
1A±1%
A±2% in the first and last quarters of the scale, ±1% in the middle half
B±3% in the first and last quarters of the scale, ±2% in the middle half

Grade A is often written "2-1-2" and grade B "3-2-3". A 10,000 psi grade A gauge may be out by ±100 psi in the middle half of its scale and ±200 psi near either end.

Choosing the accuracy you need

  1. Find the limit your job sets. A standard, a specification or a client may give one. For example, BS 8539 site tests ask for a gauge reading within 10% of the true load: see pull tester gauge accuracy.
  2. Work out the error at your lowest reading, not at full scale. Use the calculator above.
  3. Choose the range to suit. A gauge whose full scale is far above your readings carries a large error at the bottom. For dial gauges, makers following ASME B40.100 advise working between a quarter and three quarters of the scale.
  4. Check what else the data sheet states, such as an error per degree away from the reference temperature.

Test gauges and calibration

When one gauge is checked against another, the reference must be clearly more accurate than the gauge under test. A test uncertainty ratio of 4:1 is the usual aim: to check a gauge accurate to ±1% of a range, use a reference within ±0.25% of it.

Accuracy is only true while the gauge stays calibrated. A drop, an overload or plain age can shift it, which is why gauges are calibrated at least once a year. See how often to calibrate a pressure gauge and Staht's pressure gauge calibration.

Accuracy as force

On a hydraulic cylinder, the gauge's error in bar becomes an error in force. Multiply it by the cylinder's effective area: on a 25.1 cm² cylinder, ±2.5 bar is ±0.63 kN at any load. See how gauge accuracy carries into force.

Questions

Sources

  1. BS EN 837-1 Pressure gauges. Bourdon tube pressure gauges. Dimensions, metrology, requirements and testing, BSI
  2. About pressure gauge calibration: accuracy classes, % of span, Calibrate.co.uk
  3. What are ASME accuracy grades for mechanical pressure gauges?, Ashcroft
  4. How to select a pressure gauge range, Ashcroft
  5. What is the accuracy/uncertainty of the device to be tested?, Fluke
  6. Staht Heavy Duty Test Kit brochure (the G1 Pro's range and accuracy), Staht
  7. Staht G1 Pro gauge user manual, Revision 1, 04/02/2026, Staht

A gauge accurate to 0.25% of full scale